[Summary]
The standard LZ series metal pipe float flowmeters are composed of one measuring tube and on side indicator.The site indicator can be installed also with additional electric devices, suitable for liquid and gas measurement.
The flowmeter adopts to variable area measuring principle. The measuring device is composed of one taper measuring tube and floater. The floater can move freely up and down in the taper measuring pipe to change the flow area in the pipe.When the flow changes, the vertical position of the floater inside the measuring pipe changes correspondingly.Through magnetic transmitting system, the position of the floater is transmitted to the dial scale of the indicator to indicate the flow.
In M6、M8、 M9 indicator, the follow-up magnet movement is transmitted directly
to the dial scale.In M7 indicator, when the follow-up magnet rotates, the needle
driven by the cam plate indicates linearly the flow.
[Features]
[Type Selection]
Table 3 |
||||||||||||||||||||||||||||||||
LZ- |
Z |
On-site indicate |
|
|||||||||||||||||||||||||||||
|
D |
Electric transmission |
||||||||||||||||||||||||||||||
|
|
DN |
15、25、50、80、100 |
Diameter |
||||||||||||||||||||||||||||
|
|
RR1 |
Stainless steel 1Cr18Ni9Ti |
Measuring tube material |
||||||||||||||||||||||||||||
RR0 |
Stainless steel 0Cr18Ni12Mo2Ti |
|||||||||||||||||||||||||||||||
RL |
Stainless steel 00Cr17Ni12Mo2 |
|||||||||||||||||||||||||||||||
F |
PVDF or F46 fluoroplastics lining |
|||||||||||||||||||||||||||||||
|
M6 |
Pointer nonlinear indicate instant flow,LCD display instant,integrated flow |
Indicator type |
|||||||||||||||||||||||||||||
M7 |
Pointer linear indicate instant flow |
|||||||||||||||||||||||||||||||
M8 |
Multifunction indicator,explosion-proof type |
|||||||||||||||||||||||||||||||
M9 |
Pointer nonlinear indicate instant flow |
|||||||||||||||||||||||||||||||
M9B |
Stainless steel body pointer nonlinear indicate instant flow |
|||||||||||||||||||||||||||||||
|
A1 |
Battery operated no signal output,no backlight,no alarm(only for M6,M8) |
||||||||||||||||||||||||||||||
A2 |
24VDC two-wire power supply,4~20mA output(both for HART),with backlight(M9)(only for E3) |
|||||||||||||||||||||||||||||||
|
E1 |
5W1 angular displacement transmission (4~20)mA 24V DC two-wire (linear) |
Remote transmitter |
|||||||||||||||||||||||||||||
E2 |
ESK transmission with HART LCD backlight (4~20)mA 24V DC two-wire(nonlinear) |
|||||||||||||||||||||||||||||||
|
Exi |
Intrinsic safety type |
Explosion proof type |
|||||||||||||||||||||||||||||
Exd |
Explosion-proof (only for M8 indicator) |
|||||||||||||||||||||||||||||||
|
No |
No alarm |
Switch alarm output |
|||||||||||||||||||||||||||||
K1 |
With a upper switch alarm point |
|||||||||||||||||||||||||||||||
K2 |
With a floor switch alarm point |
|||||||||||||||||||||||||||||||
K3 |
With two-switch alarm point |
|||||||||||||||||||||||||||||||
|
B1 |
Vertical installation |
Structure |
|||||||||||||||||||||||||||||
B2 |
Horizontal installation |
|||||||||||||||||||||||||||||||
B3 |
Side inlet side outlet |
|||||||||||||||||||||||||||||||
B4 |
Side inlet side outlet |
|||||||||||||||||||||||||||||||
B5 |
Bottom inlet side outlet |
|||||||||||||||||||||||||||||||
|
T |
Clamp type |
Accessary |
|||||||||||||||||||||||||||||
H |
High temp.(special for LZD) |
|||||||||||||||||||||||||||||||
|
□ |
-40℃~+400℃ |
Medium temperature |
|||||||||||||||||||||||||||||
|
□ |
≤6.4MPa |
Working pressure |
|||||||||||||||||||||||||||||
|
|
□ |
g/cm³ |
Medium density |
||||||||||||||||||||||||||||
|
|
|
Q |
Gas |
Medium type |
|||||||||||||||||||||||||||
|
|
|
Y |
Liquid |
||||||||||||||||||||||||||||
LZ- |
□ |
□ |
□ |
□ |
□ |
□ |
□ |
□ |
□ |
□ |
□ |
□ |
□ |
□ |
|